PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

0 FDA warning letters since 2023 and 84 recalls in the last 12 months.

Get an email the day FDA posts anything new about PHILIPS MEDICAL SYSTEMS NEDERLAND B.V..

Recalls

ReportedClassFirmProduct and reason
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 5 M12. Model Number: 722227.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 5 M20. Model Numbers: 722228, 722281.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion S. Model Number: 722285.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7 B20. Model Numbers: 722068, 722226.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7 M12. Model Numbers: 722078, 722223.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 3 M12. Model Numbers: 722063, 722221.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7 B12. Model Numbers: 722067, 722225.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Nov 4 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 3 M15. Model Number: 722064, 722222, 722280.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Aug 19 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Interventional fluoroscopic x-ray system.
Firmware issue potentially leading to loss of geometry movements.
Aug 19 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers: R3.0 and R3.1; Interventional fluoroscopic x-ray system
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Aug 19 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 722225, 722226, 722227, 722228, 722280*, 722281*, and 722282* (*These models are distributed only i…
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Aug 12 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Allura, Cardio Vascular-Allura Centron. Model Number: 722400. Interventional Fluoroscopic X-Ray System.
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Aug 12 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; M…
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Aug 12 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper…
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Jun 24 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 70803…
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Jun 24 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 72200…
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 10 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Jun 3 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225), Azurion 7 B20 (7220…
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Jun 3 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 72201…
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
May 27 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079, 722224).
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
May 27 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 722229. 2. Azurion 3…
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
May 27 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9…
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
May 27 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (7…
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
May 20 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion…
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Apr 22 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA data.
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/15 OR Table; System Code: 722059;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 5 M12; System Code: (1)722227, (2)722231;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD10F; System Code: 722002;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/10; System Code: 722029;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/20 OR Table; System Code: 722039;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/15; System Code: 722058;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/20; System Code: 722038;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Vue Motion V12. Product Number: 1017979.
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10C; System Code: 722001;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 28 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion system; Software Version Number: R3.1;
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20; Model Number: 722028;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Product Name: Allura Xper FD10C; Model number: 722001;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number:…
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10; Model Number: 722003;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10/10; Model Number: 722011;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/20; Model Number: 722038;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10/10; Model Number: 722005;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20; Model Number: 722012;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20/10; Model Number: 722029;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numb…
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10; Model Number: 722010;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20 Biplane; Model Number: 722008;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20 OR Table; Model Number: 722023;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20 Biplane; Model Number: 722013;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD10; Model Number: 722026;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Jan 21 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Allura Xper FD20; Model Number: 722006;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Dec 3, 2025 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.